Do not operate the instrument in rooms
where risk of explosion exists!
Do not expose the instrument to direct
sunlight (windows)!

Do not install the instrument above a
heater!

3.1

Site requirements

The installation site has to meet the following
requirements:

Stable laboratory bench, exactly level, at
least 1.60 m wide and 60 cm deep.

Tap water supply at a distance of max. 2 m
and drain at a distance of max. 1.50 m from
the corresponding inlet and outlet at the
back of the instrument.

- Please bear in mind: The connections are
located at the extreme left on the back of the
instrument.

Fume hood at a distance of max. 3.50 m from
the instrument, if the instrument is to be op-
erated with the exhaust air hose (otherwise
the instrument must be operated with the
activated carbon filter)!

Vibration-free floor.

Sufficient open space (70 cm) above the lab-
oratory bench, to ensure trouble-free open-
ing of the lids.

Stable ambient temperature of +10 °C to +35
°C.

Relative air humidity of maximum 80%, non-
condensing.

No other instruments which cause vibra-
tions installed nearby.