Operation, 1 setting up the instrument, Setting up the instrument – Leica Biosystems ST4040 User Manual

Page 30

Advertising
background image

30

Instruction Manual V 1.4 – 06/2009

6.2

Preparing the instrument for staining: inserting and filling the stations

Select a staining protocol.
(For examples, see Appendix 1: ‘Staining protocols’)

According to the selected staining protocol, determine the order of re-
agent and tap water stations.

Open all three lids.

The Leica ST4040 is equipped with a three-piece lid, thus e

n

-

abling the operator to open just one of the two small lids on the right
or left of the instrument for loading / unloading slide racks. This sys-
tem keeps the exposure of laboratory personnel to hazardous fumes
to a minimum.

Remove the cover screws from those positions where you have decid-
ed to install tap water stations (tap water stations can be installed from
positions 3 through 20).

Insert the tap water stations (yellow - see left) and the reagent stations
(white).

Make sure that all stations are inserted correctly; the stations must not
be jammed (overlapping rims!).

Add reagents according to selected staining protocol. - Do not exceed
the maximum volume or remain under the minimum volume!

Close the lid entirely.

6.

Operation

6.1

Setting up the instrument

To set up the instrument follow all instructions of chapter 6 step by
step.