Leica ST4040 – Linear Stainer
4.4.2 Installing the optional unload station
on the right of the basic instrument
•
Remove the 4 hexagon socket screws from
the right lid (hexagon key, size 4).
•
Remove the lid. - Keep the screws (will be
needed later)!
•
Remove the 2 black screws from the right
side panel (hexagon key, size 3).
•
Detach the ground wire and pull it out of the
instrument (will not be needed any longer.)
4.
Installation of the Optional Load/Unload Station
•
Press the Start/Stop button, to move the
conveyor frame to the upper position. As
soon as the conveyor frame has reached the
upper end position, switch the instrument off
(ON/OFF button).
•
Remove the side panel from the side wall
and detach the ground wire from the side
wall.