2 installing the optional unload station, On the right of the basic instrument – Leica Biosystems ST4040 User Manual

Page 25

Advertising
background image

25

Leica ST4040 – Linear Stainer

4.4.2 Installing the optional unload station

on the right of the basic instrument

•

Remove the 4 hexagon socket screws from
the right lid (hexagon key, size 4).

•

Remove the lid. - Keep the screws (will be
needed later)!

•

Remove the 2 black screws from the right
side panel (hexagon key, size 3).

•

Detach the ground wire and pull it out of the
instrument (will not be needed any longer.)

4.

Installation of the Optional Load/Unload Station

•

Press the Start/Stop button, to move the
conveyor frame to the upper position. As
soon as the conveyor frame has reached the
upper end position, switch the instrument off
(ON/OFF button).

•

Remove the side panel from the side wall
and detach the ground wire from the side
wall.