Trouble shooting – Leica Biosystems ST4040 User Manual

Page 42

Advertising
background image

42

Instruction Manual V1.4 – 06/2009

Possible cause

Problem

What to do

-

Open water tap.

-

Open ball valve.

-

Call Technical Service.

-

Call in-house technical ser-
vice / plumber.

-

If bent only slightly,

read-

just carefully.

-

Reinsert correctly.

-

Parts not usable any more -
-> discard.

-

Call Technical Service.

-

Call Technical Service.

-

Install drain hose so there
is a sufficient gradient from
drain outlet on instrument
to drain pipe.

-

Remove station and clean
drain with fine bottle brush
or similar.

-

Water tap closed.

-

Ball valve closed.

-

Magnetic valve or magnetic
valve drive defective.

-

In-house plumbing
problem (pipes clogged /
furred).

-

Slide racks / slide rack car-
riers bent out of shape.

-

Slide rack carriers / slide
racks jammed.

-

Slide racks and/or slide
rack carriers significantly
bent out of shape or weld-
ing spots broken.

-

Toothed belt torn.

-

Motor or motor drive defec-
tive.

-

No or only insufficient gra-
dient of drain hose.

-

Drain of trough in basic in-
strument clogged up.

No tap water supply.

Slide racks do not advance
and/or are not lowered into re-
agent / tap water stations.

Waste water does not drain
away.

7.

Trouble Shooting

Advertising