3 staining in two rows, 1 doubling the specimen throughput, Staining in two rows – Leica Biosystems ST4040 User Manual

Page 31

Advertising
background image

31

Leica ST4040 – Linear Stainer

6.

Operation

6.3

Staining in two rows

6.3.1 Doubling the specimen throughput

•

For this type of application, the sequences of reagent / tap water sta-
tions in rows 1 and 2 must be identical.

•

According to the station sequence selected for row 1 (see Chapter 6.2)
insert an identical sequence in row 2.

•

Both rows are run with the same program.

6.3.2 Carrying out two different, but matching, staining protocols simultaneously

•

Precondition: the program parameters of both protocols must be the
same (i.e. both rows are run with the same program).

•

The two staining protocols are coordinated through the number of re-
agent stations and/or the concentration or dilution of the reagents (for
details, see Appendix 1

- ‘Staining protocols’).

Important note to chapters 6.3.1 and 6.3.2

•

When staining in two rows, standard slide rack carriers, as for single
row staining, are used.

•

Slide rack carriers located side by side in rows 1 and 2 are inserted
laterally reversed into the same notch of the conveyor frame (see left).

6.3.3 Carrying out two different staining protocols (two subsequent single loads)

•

The two rows are never carried out simultaneously, i.e.: row 1 is run
with program 1 - row 2 with program 2, or vice versa.

In 6.3.3 the advantage of staining in two rows is that 2 different staining protocols, though not
run simultaneously, can be run one directly after another without any delay; whereas in a sin-
gle load model, stations would have to be rearranged, refilled etc. prior to being able to run the
second protocol.

Advertising