Installation of the optional load/unload station – Leica Biosystems ST4040 User Manual

Page 26

Advertising
background image

26

Instruction Manual V1.4 – 06/2009

Pull the conveyor frame of the optional un-
load station upwards using both your hands.

Fit the connecting plates attached to the
front and rear half of the conveyor frame of
the optional unload station over the insides
of the front and rear half of the conveyor
frame of the basic instrument.

Place the optional unload station on the right
side of the instrument.

Move the optional unload station right up to
the instrument, inserting the two pins, locat-
ed on the left and right at the bottom of the
optional load station, into the corresponding
openings in the basic instrument.

4.

Installation of the Optional Load/Unload Station

Tighten the screw on the right wall of the op-
tional unload station (ring wrench, size10).

Open the right lid.

Plate (1) which connects optional unload
station and basic instument is fastened with
two Allan screws.

1

In order to establish safe protective
grounding, it is essential to fasten
plate (1) securely.