background image

22

Instruction Manual V1.4 – 06/2009

4.4

Installation

4.4.1 Installing the optional load station

•

Remove the 4 hexagon socket screws from
the left lid (hexagon key, size 4).

•

Remove the lid. - Keep the screws (will be
needed later)!

•

Remove the 2 black screws from the left side
panel (hexagon key, size 3).

•

Remove the side panel from the side wall
and detach the ground wire from the side
wall.

•

Detach the ground wire and pull it out of the
instrument (will not be needed any longer.)

4.

Installation of the Optional Load/Unload Station

on the left side of the basic instrument

•

Press the Start/Stop button, to move the
conveyor frame to the upper position. As
soon as the conveyor frame has reached the
upper end position, switch the instrument off
(ON/OFF button).