Installation of the optional load/unload station – Leica Biosystems ST4040 User Manual

Page 27

Advertising
background image

27

Leica ST4040 – Linear Stainer

Use the 4 hexagon socket screws which you
have kept (see step 1) to attach the lid of the
optional unload station to the corresponding
hinge (= right hinge on basic instrument).

Insert the plug of the unload sensor cable
into the unload sensor port on the rear of the
basic instrument.

To install an optional unload station on
the left side of the basic instrument,
follow the same steps - carried out lat-
erally reversed.

4.

Installation of the Optional Load/Unload Station

Lower the conveyor frame (now connected
to one single piece) pushing the ON/OFF
button.

Bolt the two front halves of the conveyor
frames together with two hexagon socket
screws (hexagon key, size 2.5).

Bolt the two rear halves of the conveyor
frames together with two hexagon socket
screws (hexagon key, size 2.5).

Advertising