Instrument properties, 1 technical data, Technical data – Leica Biosystems ST4040 User Manual

Page 29

Advertising
background image

29

Leica ST4040 – Linear Stainer

5.1

Technical Data

General

Admissions:

VDE, UL, cUL, C-Tick Label

Nominal supply voltage: available:

100 V AC ± 10 %

120 V AC ± 10 %
230 V AC ± 10 %
240 V AC ± 10 %

Nominal frequency:

50/60 Hz

Maximum power draw:

150 VA

Protective class

1

:

I

Mains fuses:

Circuit breaker manufactured by ETA, 2 A

model: 3120-F421-P7T1-W01D-2 A

Primary fuses: manufactured by Schurter: model Fst

2 x T8 A

Secondary fuses: manufactured by Schurter: model Fst

Motor (F1) T 600 mA

Electronic (F2) T 1,6 A

Pollution degree

1

:

2

Overvoltage installation category:

II

Working temperature range: +10 °C to +35 °C

Relative air humidity:

max. 80 %, non-condensing

Sound intensity level:

< 70 dB

1)

according to IEC-1010, UL 3101, EN 61010

Dimensions and weight

Basic instrument (W x H x D):

1,435 x 444 x 436 mm

Basic instrument with optional load and unload station (W x H x D):

1,969 x 444 x 436 mm

Optional load / unload station (W x H x D):

267 x 444 x 392 mm

Working height: 318 mm

Weight:

(Basic instrument with accessories)

73 kg

(optional load station)

16 kg

(optional unload station)

14 kg