Instrument properties, 1 technical data, Technical data – Leica Biosystems ST4040 User Manual
Page 29

29
Leica ST4040 – Linear Stainer
5.1
Technical Data
General
Admissions:
VDE, UL, cUL, C-Tick Label
Nominal supply voltage: available:
100 V AC ± 10 %
120 V AC ± 10 %
230 V AC ± 10 %
240 V AC ± 10 %
Nominal frequency:
50/60 Hz
Maximum power draw:
150 VA
Protective class
1
:
I
Mains fuses:
Circuit breaker manufactured by ETA, 2 A
model: 3120-F421-P7T1-W01D-2 A
Primary fuses: manufactured by Schurter: model Fst
2 x T8 A
Secondary fuses: manufactured by Schurter: model Fst
Motor (F1) T 600 mA
Electronic (F2) T 1,6 A
Pollution degree
1
:
2
Overvoltage installation category:
II
Working temperature range: +10 °C to +35 °C
Relative air humidity:
max. 80 %, non-condensing
Sound intensity level:
< 70 dB
1)
according to IEC-1010, UL 3101, EN 61010
Dimensions and weight
Basic instrument (W x H x D):
1,435 x 444 x 436 mm
Basic instrument with optional load and unload station (W x H x D):
1,969 x 444 x 436 mm
Optional load / unload station (W x H x D):
267 x 444 x 392 mm
Working height: 318 mm
Weight:
(Basic instrument with accessories)
73 kg
(optional load station)
16 kg
(optional unload station)
14 kg
5.
Instrument Properties