Installation of the optional load/unload station – Leica Biosystems ST4040 User Manual

Page 23

Advertising
background image

23

Leica ST4040 – Linear Stainer

•

Place the optional load station on the left
side of the basic instrument.

•

Pull the conveyor frame of the optional load
station upwards using both your hands.

4.

Installation of the Optional Load/Unload Station

•

Move the optional load station right up to the
instrument, inserting the two pins, located
on the left and right at the bottom of the op-
tional load station, into the corresponding
openings in the basic instrument.

•

Tighten the screw on the left wall of the op-
tional load station ring wrench, size10).

•

Plate (1) which connects optional load station
and basic instument is fastened with two Al-
lan screws.

In order to establish safe protective
grounding, it is essential to fasten plate
(1) securely.

1

•

Open the Lid