NTi Audio Acoustilyzer AL1 User Manual

Page 27

Advertising
background image

26

27

Input Range Selector & Bargraph

In the SPL/RTA Mode the AL1 has three measurement ranges (RNGE).

The span of each range depends on the sensitivity of the microphone

used. For a MiniSPL with a sensitivity of 20mV/Pa the measurement

ranges are:

20 to 100 dB 40 to 120 dB 60 to 140 dB

The selected input range is displayed at the bottom of the bargraph and

may be changed by operating the range indicator arrows. The bargraph

provides an analog display of the actual sound pressure level.

Primary Indicator Range

To achieve best measurement accuracy the appropriate range,

called the “primary indicator range”, must be selected. Correct setting

avoids overflow indications and lowers the influence of noise on the

measurement result.

The two range indicator arrows assist the user to set the optimal range.

As soon as the indicated bargraph value is found lower than the primary

indicator range, the down arrow symbol below RNGE will start to move,

symbolizing that the actual SPL measurements are inaccurate. The

bargraph range needs to be changed manually by selecting the moving

left arrow symbol and pressing the enter key.

Note: Changing the ranging resets all ongoing measurements.

When exceeding the primary indicator

range, three overload arrows replace

the big result reading.
• Select the right arrow below RNGE

and press enter to change to the next

higher bargraph range.