C. detailed test result readout – NTi Audio Acoustilyzer AL1 User Manual

Page 57

Advertising
background image

56

57

The detailed test result view of the

modulation indices and individual band

level results is provided on the second

STI-PA measurement screen.

c. Detailed Test Result Readout

STI-PA

Octave Bands

Individual Octave

Band LEQ Level

STI-PA Modulation

Index 2

STI-PA Modulation

Index 1

Error

Detection

STI-PA Modulation Index 1 & 2:

For high speech intelligibility it is mandatory that the integrity of the

transmitted voice signal modulations are preserved. Therefore STI-PA is

based on measuring the MTF (Modulation Transfer Function). This function

quantifies the degree to which the voice modulations are preserved in

individual octave bands. STI-PA determines the MTF by analyzing all

seven frequency bands, each band being modulated with two frequencies,

resulting in the modulation index 1 and index 2