Calibration – NTi Audio Acoustilyzer AL1 User Manual

Page 62

Advertising
background image

62

63

Calibration

In order to obtain correct measurement results, the sensitivity of the micro-

phone connected to the AL1 must be set and calibrated accordingly. The

microphone sensitivity is set by default to the NTi Audio MiniSPL measure-

ment microphone factory adjustment 20mV/Pa. MiniSPL is the standard

measurement microphone for the Acoustilyzer AL1. Together they form a

comprehensive class 2 acoustical analyzer.

Alternatively to the MiniSPL, the NTi Audio M2010 class 1 frequency response

microphone can be used for highly accurate class 1 measurements.

Microphone

Sensitivity

Microphone Balance Indicator,

arrow has to be centered to verify

the balanced mic input singal

Default

Factory Setting

Manual Calibration with

external microphone calibrator

The following three calibration modes are available:

DEFAULT

All MiniSPL microphones are factory

calibrated to a sensitivity of 20.0 mV/Pa.

Reset to default setting by:

• Select DEFAULT
• Press the enter key