Nleq – NTi Audio Acoustilyzer AL1 User Manual

Page 33

Advertising
background image

32

33

⎟⎟

⎜⎜

=

=

N

n

dt

LEQ

n

N

LEQ

1

10

_

10

1

log

10

MIN_dt Minimum SPL value measured within the actual log interval

∆t;

time weighted value (SLOW/FAST).

MAX_dt Maximum SPL value measured within the actual log interval

∆t;

time weighted value (SLOW/FAST).

Example of an SPL/LEQ log visualized with Microsoft Excel

(without LCpeak)

LCpeak Maximum C-weighted peak sound pressure level during the

whole measurement period. The measurement period is

indicated and controlled by the measurement timer.

LCpeak_dt Maximum C-weighted peak sound pressure level during the

log interval

∆t

.

SPL_Act Indicates the actual SPL level measured exactly at the end

of the log interval. Note that Leq_dt, MIN_dt and MAX_dt

characterize an interval more appropriately than this single

SPL_Act value.