NTi Audio Acoustilyzer AL1 User Manual

Page 38

Advertising
background image

b. How to measure

The room under test is injected with a pink noise signal through speakers.

The sound source has to be active until a balance between injected and

absorbed acoustical energy has been reached. i.e. allow enough time for

the sound to reflect off all surfaces. As a rule of thumb, the noise should

be played for at least the number of seconds recorded in the RT60 result.

The source signal is then stopped. The AL1 recognizes this interruption,

triggers, the decay time is measured and the RT60 test result is automatically

calculated.

Please follow the RT60 test instruction below:

Ranging
• Select RANGE and press the enter key

to measure the actual sound pressure

level in the room without a test signal

present.

• The SET RANGE field pops up. Confirm

the automatic ranging with OK.

Now the RT60 upper mark indicators are

set to the minimum pink noise level needed

to successfully complete a measurement.

The lower marks indicate the environmental

noise without any test signal present.