e. STI-PA Error Detection

The Acoustilyzer STI-PA measurement function has a built in error detection

which helps you identifying faulty measurements. The error detection

checks the following parameters:

Invalid modulation indices (MF1 or MF2 > 1.3)

Irregularities during ongoing measurements

Both parameters depend on the amount of impulsive environmental noise.

Impulsive environmental noise influences the accuracy of any STI-PA

measurement.

If the error detection finds any problem,

the common STI-PA test result display is

interrupted with flashing “?.??” question

marks.

This error indication might be caused either

by
• missing test signal level, or
• impulsive background noise events

Detailed test result screen

The octave bands detected with

irregularities, are marked with a question

mark (?).

Recommendation

If the error detection finds any faults,

NTi Audio recommends you repeat the

measurement and compare the STI test

result with the previous ones.