RKI Instruments SDM-E2 User Manual

Page 24

Advertising
background image

19 • Setting the Operational Parameters in Edit Mode

Setting the Calibration Parameters

Do the following to set the calibration parameters after establishing a connection
between an EAGLE 2 and the SDM-E2.

1.

Turn on the SDM-E2 with an EAGLE 2 and establish a connection between
them as described above in "Turning on the SDM-E2 with an EAGLE 2". The
EAGLE 2 will display the home screen.

2.

Press and hold the EDIT ENTER button for 2 seconds. The EAGLE 2 will
display the Edit Mode screen.

3.

Press and release the CAL

S

button. The EAGLE 2 will display the following

screen with the two calibration parameters and their settings.

CH4 50%LEL GAS1

OXY 12.0vol% GAS1

H2S 25.0ppm GAS1

CO 50ppm GAS1

---
---

EDIT MODE

BUMP TEST : BUMP

CALIBRATION : CAL

GAS INLET : EDIT

CAL PARAMETER

1.AIR TIME: 30sec
2.CAL TIME: 60 sec