39 • Calibrating Instruments with Standard Sensors

13. The SDM-E2 will perform a fresh air adjustment on the instrument.

If the air adjustment is successful, the SDM-E2 will continue as described
in step 14 below.

If one or more of the sensors fails the fresh air adjustment, the SDM-E2 will
abort the calibration and will not apply calibration gas. If this happens, the
CAL

S

LED will flash red indicating a failure and the following screen will

appear.

In this case continue with step 17.

14. The SDM-E2 will apply calibration gas to the instrument for the time defined by

the GAS TIME calibration parameter.

The “C” indicates calibration gas is being applied.

15. The SDM-E2 will then purge the system with fresh air for the time defined by

the AIR TIME calibration parameter. The calibration results will be displayed

CH4 0%LEL Z P

OXY 20.9vol% Z P

H2S 5.0ppm Z P

CO 0ppm Z P

---
---

CH4 0%LEL Z P

OXY 20.9vol% Z P

H2S 25.0ppm Z F

CO 0ppm Z P

---
---

CH4 29%LEL C

OXY 18.0vol% C

H2S 25.0ppm C

CO 45ppm C

---
---