RKI Instruments SDM-E2 User Manual

Page 37

Advertising
background image

Bump Testing Instruments with Standard Sensors • 32

15. The SDM-E2 will apply calibration gas to the instrument for the time defined by

the GAS TIME bump test parameter. A “B” will appear to the right of the gas
reading to indicate that calibration gas is being applied for a bump test.

At the end of the calibration gas application, the SDM-E2 will determine if the
bump test passed.

16. When the AUTO CAL bump test parameter is set to OFF:

a. The instrument will indicate which channels passed or failed the

bump test with a P (pass) or an F (fail) to the right of the “B” which
indicates it is a bump test result. The bump test gas readings at the
end of the gas application will be displayed.

b. The BUMP

T

LED will continue to flash amber while the docking

station performs a fresh air purge for the time defined by the AIR
TIME bump test parameter. After the purge is complete, if all
channels passed the bump test, the BUMP

T

LED will turn solid

green. If any channels failed the bump test, the BUMP

T

LED will

turn solid red.

c. Continue to step 18.

CH4 30%LEL B

OXY 9.0vol% B

H2S 3.0ppm B

CO 15ppm B

---
---

CH4 48%LEL BP

OXY 12.0vol% BP

H2S 3.0ppm BF

CO 47ppm BP

---
---