Bump Testing an Instrument with Special Sensors • 50

e. Once the GAS 2 calibration gas has been applied for the

appropriate amount of time, the SDM-E2 will determine if the sixth
channel passed the bump test and will indicate a P (pass) or an F
(fail) to the right of the “B” which indicates it is a bump test result. If it
passed, the following screen will appear.

f.

The BUMP

T

LED will continue to flash amber while the docking

station performs a fresh air purge for the time defined by the AIR
TIME bump test parameter. After the purge is complete, The BUMP

T

LED will turn solid green.

g. Continue to step 18.

If any of the standard four channels fails the bump test:

a. A calibration is automatically started on the standard four channels

and the GAS 1 calibration gas will continue to be applied. The
current gas readings will be displayed along with the bump test
results. A “C” will appear next to the channels indicating they are
being calibrated.

CH4 48%LEL BP

OXY 12.0vol% BP

H2S 23.0ppm BP

CO 47ppm BP

CO2 4500ppm BP

NH3 23.0ppm BP

CH4 29%LEL BF C

OXY 18.0vol% BF C

H2S 25.0ppm BP C

CO 45ppm BF C

CO2 ---------

NH3 ---------