Bump Testing an Instrument with Special Sensors • 54

instrument, it will automatically turn it off 10 minutes after the last button
push.

To turn off the instrument before it is automatically turned off, press and
hold the EDIT ENTER and POWER buttons simultaneously for at least one
second and then release them.

CAUTION:

When using the EAGLE 2 with the SDM-E2, do not turn off the
instrument using the instrument power button. Use the EDIT ENTER
and POWER buttons on the SDM-E2 to turn off the instrument.

The BUMP

T

or BUMP

T

and CAL

S

LEDs will remain on indicating the

test results. If the same EAGLE 2 is turned on again, the test results will still
be indicated by the BUMP

T

or BUMP

T

and CAL

S

LEDs and on the

EAGLE 2 screen. To clear the BUMP

T

or BUMP

T

and CAL

S

LEDs,

with the EAGLE 2 on, press and release the EDIT ENTER button to return
to the home screen. If a new EAGLE 2 is turned on and connected, the
results displayed by the BUMP

T

or BUMP

T

and CAL

S

LEDs will

automatically be cleared.

19. The results of the bump test or bump test and calibration will be stored in the

SDM-E2’s memory and will be available to copy to a USB flash drive. See
“Copying Calibration and Bump Test Records to a USB Flash Drive” on
page 68 for instructions to copy the saved bump test and calibration records to
a USB flash drive.

20. Disconnect the charging cable, if connected, and the gas out to EAGLE 2 line

from the EAGLE 2.

21. Remove the EAGLE 2 from the SDM-E2.

22. If you wish to bump test additional instruments, repeat step 5 - step 21 above

for each additional instrument.