RKI Instruments SDM-E2 User Manual

Page 41

Advertising
background image

Calibrating Instruments with Standard Sensors • 36

To perform a calibration on an instrument:

1.

Confirm that the AC Adapter is connected to the SDM-E2 and to an AC wall
socket.

2.

Press and hold the SDM-E2’s POWER button. The LEDs will turn amber.

3.

When the BUMP

T

and CAL

S

LEDs turn off, release the POWER button.

4.

The COPY LED will be off or on steadily and the CHARGE LED will be blinking
green if the SDM-E2 is operating properly or solid red if there is a system
failure. The amount of free memory in the SDM-E2 will dictate the condition of
the COPY LED (see “Available Memory in the SDM-E2” on page 68).

5.

Install an EAGLE 2 in the instrument cradle. Set the EAGLE 2 onto the top of
the SDM-E2 and slide the instrument forward until it falls into place as
illustrated below.

Figure 12: Inserting the EAGLE 2

WARNING:

Inserting the EAGLE 2 improperly may damage the exhaust bellow at the
back of the SDM-E2.