RKI Instruments SDM-E2 User Manual

Page 36

Advertising
background image

31 • Bump Testing Instruments with Standard Sensors

14. The SDM-E2 will perform a fresh air adjustment on the instrument.

If the air adjustment is successful, the SDM-E2 will indicate that all
channels passed with a “P” to the right of the “Z”. The “Z” indicates it is a
fresh air adjustment (zero) result. The SDM-E2 will continue as described
in step 15 below.

If one or more of the sensors fails the fresh air adjustment, the SDM-E2 will
abort the bump test and will not apply calibration gas. If this happens, the
BUMP

T

LED will flash red indicating a failure and the following screen will

appear indicating which channels passed and failed the fresh air
adjustment with a P (pass) or an F (fail) to the right of the “Z”. The “Z”
indicates it is a fresh air adjustment (zero) result. In the following example,
the H

2

S channel failed the zero adjustment.

In this case, continue with step 18.

CH4 0%LEL Z P

OXY 20.9vol% Z P

H2S 0.0ppm Z P

CO 0ppm Z P

---
---

CH4 0%LEL Z P

OXY 20.9vol% Z P

H2S 5.0ppm Z F

CO 0ppm Z P

---
---

Advertising