RKI Instruments SDM-E2 User Manual

Page 54

Advertising
background image

49 • Bump Testing an Instrument with Special Sensors

b. GAS 2 calibration gas will be applied to the fifth sensor and the

SDM-E2 will determine if it passed bump testing. If it passed, the
following screen will appear:

NOTE:

If your instrument has only one special sensor, continue to step 18.

c. If your instrument has more than one special sensor, as in this

example, you will need to change the GAS 2 cylinder in order for the
bump test to continue. In the example below, the EAGLE 2’s second
special sensor is an ESM-01 for NH

3

.The EAGLE 2 screen will

appear as shown below:

d. Once you have changed the GAS 2 cylinder to the appropriate gas,

press and release the EDIT ENTER button on the SDM-E2. The
bump test will continue.

CH4 48%LEL BP

OXY 12.0vol% BP

H2S 23.0ppm BP

CO 47ppm BP

CO2 4500ppm BP

NH3 ---------

CHANGE GAS2

TO NH3

OK : EDIT

CH4 48%LEL BP

OXY 12.0vol% BP

H2S 23.0ppm BP

CO 47ppm BP

CO2 4500ppm BP

NH3 3.0ppm B