Setting the Operational Parameters in Edit Mode • 20

4.

If you wish to cancel setting the calibration parameters or were just viewing the
parameters to confirm their values, press and release the CAL

S

button to

return to the home screen.

To continue, press and release the EDIT ENTER button. A cursor will appear
to the right of the first calibration parameter. For a description of the calibration
parameters, see “Bump Test & Calibration Parameters” on page 13.

5.

Use the BUMP

T

and CAL

S

buttons to move the cursor up and down until it

is next to the desired parameter. Do not move the cursor down past the CAL
TIME selection unless you would like to return to the home screen.

6.

With the cursor next to the desired parameter, press and release the EDIT
ENTER button. An asterisk (*) will appear on the EAGLE 2 display to the left of
the cursor indicating that the parameter value can be changed. In the example
below, the AIR TIME parameter has been selected for updating.

7.

Use the BUMP

T

and CAL

S

buttons to set the parameter to the desired

value, then press and release the EDIT ENTER button. The asterisk next to
the cursor will disappear.

8.

Repeat step 5 - step 7 to set any other parameters.

9.

When you are done setting the parameters, use the BUMP

T

button to move

the cursor down past CAL TIME. The screen will indicate that the parameter
changes have been saved and the EAGLE 2 will return to the home screen.

CAL PARAMETER

1.AIR TIME: 30sec <

2.CAL TIME: 60 sec

CAL PARAMETER

1.AIR TIME: 30sec *<

2.CAL TIME: 60 sec