67 • Charging an Instrument in a Docking Station

Batteries Too Drained for Standalone Operation

If the charger is connected to the EAGLE 2 and the batteries in the EAGLE 2 do not
have enough power to keep the instrument on, it will go into a 20 minute charge cycle
and the following screen will be displayed on the EAGLE 2.

It will not connect to the SDM-E2. The instrument cannot be bump tested or
calibrated in the charge cycle. Once the 20 minute charge cycle has ended, the unit
will automatically turn back on and connect to the SDM-E2. The unit can now be
bump tested or calibrated.

LOW BATTERY

PLEASE WAIT

WHILE CHARGING

19’59”