RKI Instruments SDM-E2 User Manual
Page 57

Bump Testing an Instrument with Special Sensors • 52
d. Once the calibration is complete, the screen will indicate whether
the fifth channel calibration passed or failed.
NOTE:
If your instrument has only one special sensor, continue to step i.
e. If your instrument has more than one special sensor, as in this
example, you will need to change the GAS 2 cylinder in order for the
bump test to continue. In the example below, the EAGLE 2’s second
special sensor is an ESM-01 for NH
3
. The EAGLE 2 screen will
appear as shown below:
f.
Once you have changed the GAS 2 cylinder to the appropriate gas,
press and release the EDIT ENTER button on the SDM-E2. The
bump test will continue.
CH4 29%LEL BF CF
OXY 18.0vol% BF CF
H2S 25.0ppm BP CP
CO 45ppm BF CP
CO2 4500ppm BF CP
NH3 ---------
CHANGE GAS2
TO NH3
OK : EDIT
CH4 29%LEL BF CF
OXY 18.0vol% BF CF
H2S 25.0ppm BP CP
CO 45ppm BF CP
CO2 4500ppm BF CP
NH3 3.0ppm B