Bump Testing an Instrument with Special Sensors • 52

d. Once the calibration is complete, the screen will indicate whether

the fifth channel calibration passed or failed.

NOTE:

If your instrument has only one special sensor, continue to step i.

e. If your instrument has more than one special sensor, as in this

example, you will need to change the GAS 2 cylinder in order for the
bump test to continue. In the example below, the EAGLE 2’s second
special sensor is an ESM-01 for NH

3

. The EAGLE 2 screen will

appear as shown below:

f.

Once you have changed the GAS 2 cylinder to the appropriate gas,
press and release the EDIT ENTER button on the SDM-E2. The
bump test will continue.

CH4 29%LEL BF CF

OXY 18.0vol% BF CF

H2S 25.0ppm BP CP

CO 45ppm BF CP

CO2 4500ppm BF CP

NH3 ---------

CHANGE GAS2

TO NH3

OK : EDIT

CH4 29%LEL BF CF

OXY 18.0vol% BF CF

H2S 25.0ppm BP CP

CO 45ppm BF CP

CO2 4500ppm BF CP

NH3 3.0ppm B