Calibrating an Instrument with Special Sensors • 62

EDIT ENTER button until the home screen appears.

To turn the EAGLE 2 off:

If the calibration was successful, the instrument will shut off after 15
seconds. If the calibration failed, the instrument will shut off after 10
minutes. If buttons are pressed before the SDM-E2 turns off the
instrument, it will automatically turn it off 10 minutes after the last button
push.

To turn off the instrument before it is automatically turned off, press and
hold the EDIT ENTER and POWER buttons simultaneously for at least one
second and then release them.

CAUTION:

When using the EAGLE 2 with the SDM-E2, do not turn off the
instrument using the instrument power button. Use the EDIT ENTER
and POWER buttons on the SDM-E2 to turn off the instrument.

The CAL

S

LED will remain on indicating the test results. If the same

EAGLE 2 is turned on again, the test results will still be indicated by the
CAL

S

LED and on the EAGLE 2 screen. To clear the CAL

S

LED, with

the EAGLE 2 on, press and release the EDIT ENTER button to return to
the home screen. If a new EAGLE 2 is turned on and connected, the
results displayed by the CAL

S

LED will automatically be cleared.

23. The results of the calibration will be stored in the SDM-E2’s memory and will

be available to copy to a USB flash drive. See “Copying Calibration and Bump
Test Records to a USB Flash Drive” on page 68 for instructions to copy the
saved bump test and calibration records to a USB flash drive.

24. Disconnect the charger cable, if connected, and the gas out to EAGLE 2 line

from the EAGLE 2.

25. Remove the EAGLE 2 from the SDM-E2.

26. If you wish to calibrate additional instruments, repeat step 5 - step 21 above

for each additional instrument.