Bump Testing Instruments with Standard Sensors • 34

The current gas readings will be displayed along with the bump test
results. A “C” will appear next to the channels indicating they are
being calibrated. The BUMP

T

LED will continue to flash amber and

the CAL

S

LED will begin flashing amber to indicate a calibration is

taking place.

b. At the end of the calibration, the instrument displays the calibration

gas readings at the end of the gas application and the results from
both the bump test and the calibration with a P (pass) or an F (fail).

c. The BUMP

T

LED and CAL

S

LED will continue to flash amber

while the docking station performs a fresh air purge for the time
defined by the AIR TIME bump test parameter. After the purge is
complete, the BUMP

T

LED will turn solid red. If all channels

passed the calibration, the CAL

S

LED will turn solid green. If any

channels failed the calibration, the CAL

S

LED will turn solid red.

18. After a successful or failed bump test,

To perform any other operations:

To perform another bump test, press and hold the BUMP

T

button until the

pump starts. To perform a calibration, press and hold the CAL

S

button

until the pump starts. To return to the home screen, press and hold the
EDIT ENTER button until the home screen appears.

To turn the EAGLE 2 off:

If the bump test was successful, the instrument will shut off after 15