RKI Instruments SDM-E2 User Manual

Page 27

Advertising
background image

Setting the Operational Parameters in Edit Mode • 22

3.

Press and release the EDIT ENTER button. The EAGLE 2 will display the
following screen with the two gas inlet parameters and a cursor next to the first
parameter. The USING setting will be visible.

4.

To change the USING setting, move the cursor next to the USING parameter
with the BUMP

T

or CAL

S

button if necessary and press and release the

EDIT ENTER button. An asterisk will appear to the left of the cursor.

5.

Use the BUMP

T

and CAL

S

buttons to scroll through the options: GAS1

ONLY, GAS1&GAS2, and GAS2 ONLY. If GAS1 ONLY is selected, when a
bump test or calibration is performed, only those channels assigned to GAS 1
will be bump tested or calibrated. Similarly, if GAS2 ONLY is selected, when a
bump test or calibration is performed, only those channels assigned to GAS 2
will be bump tested or calibrated. If GAS1&GAS2 is selected, all active
channels will be bump tested or calibrated when one of these operations is
performed.

6.

Once you have made your selection, press and release the EDIT ENTER
button. The asterisk will disappear.

GAS INLET PARAMETER

1.USING : GAS1&GAS2 <
2.ASSIGN : . . .



GAS INLET PARAMETER
1.USING : GAS1&GAS2*<

2.ASSIGN : . . .