29 • Bump Testing Instruments with Standard Sensors

WARNING:

Do not plug the charger cable into a battery pack that contains alkaline
batteries. Do not attempt to charge alkaline batteries.

7.

Press and hold the POWER ENTER RESET button on the EAGLE 2 until you
hear a beep, then release it. The EAGLE 2 will begin its power up sequence. If
a successful connection between the EAGLE 2 and the SDM-E2 occurs, the
home screen will display at the end of the start up sequence. If the charge LED
was amber, it will begin to blink green.

NOTE:

The screen shown above applies to a standard 4-channel EAGLE 2. If
your EAGLE 2 has less than 4 channels, the inactive channels will not
appear in the above screen or in any screens where channels are
displayed. If your EAGLE 2 has a different gas combination, your gas
names and auto calibration values will be different. All screens in this
section assume a standard 4-channel EAGLE 2.

8.

If necessary, confirm that the bump test check tolerance is set to the desired
value. See “Setting the Bump Test Parameters” on page 17.

9.

Verify that the appropriate calibration gas cylinder is connected to the GAS 1
and GAS 2 fittings on the back of the SDM-E2. See “Connecting Calibration
Gas” on page 24 for calibration gas cylinder options and calibration gas
connection procedures.

10. Connect the gas out to EAGLE 2 line to the inlet fitting on the EAGLE 2.

CH4 50%LEL GAS1

OXY 12.0vol% GAS1

H2S 25.0ppm GAS1

CO 50ppm GAS1

---
---