Bump Testing an Instrument with Special Sensors • 46

15. The SDM-E2 will apply GAS 1 calibration gas to the instrument for the time

defined by the GAS TIME bump test parameter. The fifth and sixth

channels’

gas reading will be replaced by dashes while the standard 4 sensors are being
bump tested.

At the end of the calibration gas application, the SDM-E2 will determine if the
standard 4 sensors passed the bump test.

16. When the AUTO CAL bump test parameter is set to OFF:

a. The instrument will indicate which standard 4 channels passed or

failed the bump test with a P (pass) or an F (fail) to the right of the
“B” which indicates it is a bump test result. The bump test gas
readings at the end of the gas application will be displayed. The fifth
channel will then be bump tested with the GAS 2 cylinder.

CH4 48%LEL B

OXY 12.0vol% B

H2S 3.0ppm B

CO 47ppm B

CO2 ---------

NH3 ---------

CH4 48%LEL BP

OXY 12.0vol% BP

H2S 3.0ppm BF

CO 47ppm BP

CO2 3500ppm B

NH3 ---------