3 meas tc cond - measuring parameters – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1045

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1029

5.6.4.11.3

MEAS TC Cond - Measuring parameters

Tab: Method

▶ MEAS TC Cond ▶ Properties... ▶ Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.

Measurement frequency

712
Selection

auto | 300 Hz | 2.4 kHz

Default value

auto

auto
The more suitable of the two frequencies is selected.

Stop time

Maximum period of time for measurement.

Range

0 ... 999999 s

Default value

3000 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

712
Range

0.4 ... 999999 s

Default value

2.0 s

856
Range

0.1 ... 999999 s

Default value

2.0 s

Sample solution

Selection of the sample solution from the table Sample solutions (TC
conductivity)
.

Selection Table entries
Default value first entry of table

Start temperature

Temperature at the start of the measurement.

Range

-20.0 ... 150.0 °C

Default value

20.0 °C