Metrohm tiamo 2.3 (process analysis) User Manual

Page 844

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

828

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead.

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

'Device types' | Titrando

Default value

Titrando

Sensor

Sensor

Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode

I(pol)

Polarization current at the indicator electrode.

Selection

2 µA | 5 µA | 10 µA | 20 µA | 30 µA

Default value

10 µA

Note

When changing this parameter, make sure that the polarization current
of the indicator electrode, the endpoint and the control range are
linked closely with one another (see Chapter 5.6.3.6.4, page 831).

Electrode check

on | off (Default value: on)

If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.

Temperature measurement

Type of temperature measurement.

Advertising