5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1154

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Dosing device

Selection of the second dosing device (exchange or dosing unit) to be
used for the dosing if the first dosing device is not available.

Titrando, 846, 856, 867

Selection

1 | 2 | 3 | 4

Default value

2

814, 815, 855, 864, 874

Selection

1 | 2 | 3

Default value

2

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked.

Titrando, 846, 856, 867, 814, 815, 855, 864, 874
Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Filling rate

Speed with which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the
exchange unit or dosing device used. If the entered filling rate is too high
for the selected dosing device, it will automatically be reduced during dos-
ing to the largest possible value.

Titrando, 846, 856, 867, 814, 815, 855, 864, 874
Range

0.01 ... 166.00 mL/min

Default value

maximum mL/min

Stirrer

Stirrer

Selection of the stirrer.

Titrando, 846, 856, 867, 814, 815, 855, 864, 874
Selection

1 | 2 | 3 | 4 | off

Default value

1