3 properties - calibration data – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1435

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1419

6.7.3.3

Properties - Calibration data

Tab: Configuration

Sensors Edit Properties Calibration data

Note

This tab is displayed only for sensors of the type pH electrode, ISE
electrode
, conductivity sensor and Thermoprobe.

Specific calibration data for pH electrodes

Slope

Slope of the pH electrode. This value can be determined automatically
with a calibration from the linearized calibration curve or entered man-
ually. The slope of a pH electrode is available for calculations as the varia-
ble 'Command name.SLO.

Range

-999.9 ... 999.9 %

Default value

100.0 %

pH(0)

After the slope, pH(0) is the second characteristic of the calibration func-
tion U = f(pH). pH(0) is the x-axis intercept of the calibration curve, i.e. it
corresponds to the pH value at 0 mV. The electrode zero point pH(0) of a
pH electrode is available as a variable 'Command name.ENP' for calcu-
lations.

Range

-20.000 ... 20.000

Default value

7.000

Specific calibration data for ISE electrodes

Slope

Slope of the ISE electrode. This value can be determined automatically
with a calibration from the linearized calibration curve or entered man-
ually. The slope of an ISE electrode is available for calculations as the varia-
ble 'Command name.SLO.

Range

-999.9 ... 999.9 mV

Default value

59.2/Valence mV

E(0)

E(0) is the second characteristic of the calibration function U = f(log c).
E(0) is the y-axis intercept of the calibration curve and corresponds to the
potential at log c = 0. The electrode zero point E(0) of an ISE electrode is
available as a variable 'Command name.ENP' for calculations.

Advertising