4 meas ph - evaluations – Metrohm tiamo 2.3 (process analysis) User Manual

Page 957

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

941

Stop measured value pH

Stops when the preset measured value has been reached since the start of
the measurement.

Titrando, 855, 867
Range

-20.000 ... 20.000

Selection

off

Default value

off

off
No stop.

Temperature

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.

Titrando, 855, 867
Range

-20.0 ... 150 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500 °C

Default value

25.0 °C

5.6.4.3.4

MEAS pH - Evaluations

Tab: Method

MEAS pH Properties... Evaluations

Command name

Name of the command.

Input

25 characters

The following methods for evaluation of measurement curves can be acti-
vated and defined on this tab.

Fixed endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fix measured quantity (Meas-
ured value
or Time) for the fix endpoint from the measuring point list.

Advertising