Metrohm tiamo 2.3 (process analysis) User Manual

Page 935

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

919

Selection

off

Default value

off

Stop slope

Stop when the criterion defined for the first derivative of the measured
value (slope) is fulfilled.

First field: Selection of the reference condition.

Selection

< | >

Default value

<

Second field: Definition of reference value for the derivative of the meas-
ured value.

Range

-999999.999 ... 999999.999 °C/mL (Increment:
0.001)

Default value

0.100 °C/mL

Selection

off

Stop slope active after

Required minimum volume for the stop. The field Stop slope active
after
is active only if the comparison value is not set to off for Stop
slope
.

Range

0.00 ... 50.00 mL

Selection

off

Default value

off

Additional volume after stop

After stopping the titration, this additional volume is dosed according to
the stop conditions defined. The curve shape after reaching the endpoint
can also be seen this way.

Range

0.01 ... 50.00 mL

Default value

0.5 mL

Selection

off

Filling rate

Rate (volume/time unit) at which the buret is to be filled after the titration.
The maximum filling rate depends on the cylinder volume of the dosing
unit used. If the selected filling rate is higher than the maximum possible
dosing rate, it will be automatically corrected to this value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum