5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

891

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Input

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Tandem dosing

on | off (Default value: off)

If this option is enabled, the dosing is carried out without interruption
with a combination of two dosing devices so that the second dosing
device is dosing while the first one is being filled and vice-versa (see Chap-
ter 5.6.3.8.5, page 859)
. This parameter is displayed only for Titrandos
and 855.

Dosing device

Selection of the number of the dosing device (dosing or exchange unit) to
be used for the dosing. All the dosing device connections which are possi-
ble with the selected device type are displayed.

Titrando
Selection

1 | 2 | 3 | 4

Default value

2

855
Selection

1 | 2 | 3

Default value

2

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method run to see whether the
correct solution has been set on the selected dosing device and whether
the dosing device type is correct. With non-intelligent exchange or dosing
units, only the cylinder volume is checked.