5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

869

off
The measured vales will not be applied until after the maximum wait-
ing time has passed.

Min. waiting time

The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.

Range

0 ... 999999 s

Default value

1 s

Pause 1

Pause 1

Waiting time, e.g. for the electrode to settle down before a start volume is
added.

Titrando, 855
Range

0 ... 999999 s

Default value

0 s

Start volume

Start volume

Volume to be added before the start of the titration at the dosing rate
indicated.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL

Titrino
Range

0.00 ... 999.99 mL

Default value

0.00 mL