3 brc - start conditions – Metrohm tiamo 2.3 Manual User Manual

Page 862

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

846

■■■■■■■■

tiamo 2.3

Switch on/off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched on automatically at
the start of the command and will be switched off automatically at the
end of the command.

5.6.3.7.3

BRC - Start conditions

Tab: Method

BRC Properties... Start conditions

Command name

Name of the command.

Input

25 characters

Initial measured value

Note

Only displayed for Titrando and 855.

The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:

Measured value drift

The measured value is only accepted if the drift is less than the value
entered here.

Range

0.1 ... 999.0 mV/min

Selection

off

Default value

off

off
The measured vales will not be applied until after the maximum wait-
ing time has passed.

Min. waiting time

The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Advertising