ULTRA-CUT 400 XT

Manual 0-5306

APPENDIX

A-73

No Power to the Pump Motor

The pump motor is powered by 230 VAC controlled by the MC3 control relay. During the 4 minutes after turning on power,

before the 404 fault code is displayed, measure for 230 VAC at the motor connector J16 pin 1 to pin 3.

a. If the pump motor is not getting 230 VAC, measure for 24 VAC on the coil of MC3. If present and the relay contacts aren’t

closed the relay is defective. Note, the coil is rectified so you won’t measure continuity of even a good coil.

b. If 24 VAC is not on the MC3 coil check for D27 on the relay board being ON. If it’s on the Relay board should be providing

the 24 VAC so if it’s not the Relay board may be defective. Measure for 24 VAC at J9-6 to J9-12 on the Relay board. If

24 VAC is present and D27 is on, the Relay board or the wire harness is defective.

c. If D27 is not on, measure on the CCM I/O board between TP3 and the common at TP1. It should be low, near zero volts.

If not the CCM is probably defective. Jumper TP3 (I/O board) to TP1. If the pump comes on now replace the CCM.

d. If jumping TP3 to TP1 does not turn the fans on then the Relay board or the 40 pin ribbon cable pin 13 is at fault.

Art # 12313

MC3A

MC3B

M1

Torch Coolant Pump

1

2

3

J16

1

2

3

4

J13

(65B)

(66)

(67)

(64B)

CHASSIS GND

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

J8

1

5

3

2

4

Coolant Pump Control

K5

MC3

Coolant pump Control

(163)

(162)

SA4

ARC_SUPPRESSOR

D27

24 VAC

Relay PCB

TP1

TP3

CCM I/O PCB

To test Pump relay jump TP3 to TP1.

+24

J4-13

230 VAC from T1

Coolant flows but flow is less than the required minimum:

Test and adjust the pump/bypass valve:

This test measures the “dead head” or blocked flow pressure at the rear panel coolant supply fitting. Perform this test only

after the coolant system is fully primed, that is after the coolant is circulated throughout the system and is mostly free of

bubbles. It requires a pressure gauge with #6 JIC fitting.

The gauge needs to be able to read at least 173 PSI. Remove the coolant supply hose and connect the pressure gauge in

its place. For the Auto-Cut 200 XT connect the gauge in place of the torch coolant supply hose on the torch connection

bulkhead. This is a #5 JIC fitting.

NOTE:

Do not put the gauge in-line and attempt to pinch off the hose to block the flow. It is very difficult to totally block the

flow and failure to do so will result in incorrect setting of the bypass.

Turn the on the unit. You will have 4 minutes to perform the test/adjustment before the system times out with a coolant flow

fault. If that happens recycling the power gives you another 4 minutes.

1. The pressure on the gauge should be close to 173 (170-175) PSI. If it is the pump and the bypass are OK.

2. If the pressure is less than 173 PSI adjust the bypass screw clockwise to raise the pressure. If you can change the

pressure with the bypass screw but cannot reach 173 PSI it is likely the pump is worn out or damaged. If the pres-

sure does not change using the bypass screw it is likely the bypass is defective.

3. If the pressure is above 173 reduce the pressure by adjusting the bypass valve screw counter clockwise.