BATTERY/AC OPERATION

The LTC3000we/wi may be operated on battery power or using a USB

computer/ wall adapter. If using the meter as a bench top unit, use the

AC wall adapter if possible to extend the battery life.
To charge the battery with the wall adapter, plug the smaller end of the

USB cable (USB mini B connector) into the meter and the larger end of

the USB cable (USB Type A connector) into the wall adapter. Plug the

wall adapter into an AC outlet.
To charge the battery from a computer, plug the smaller end of the USB

cable (USB mini B connector) into the meter and the larger end of the

USB cable (USB Type A connector) into a USB port on the computer..
The battery icon will show no bars and fl ash when the unit fi rst turns on.

Then the indicator will indicate the battery status by showing 0, 1, 2, 3 or

4 bars.
It will take 5 hours to fully charge a low battery. The battery icon will

fl ash when the battery is charging. The battery icon will show four bars

and stop fl ashing when it is fully charged. The charging circuit will

automatically switch to a fl oat charge when the battery is fully charged.

The charger may remain connected. Some computers will NOT supply

power to their USB ports during standby operation. The wall charger will

charge the unit continuously.
The battery icon will show no bars and continuously fl ash if the battery

is getting low but the unit will still operate normally. A “Low Battery”

message on the status bar of the display will replace the time when the

battery voltage is too low for proper operation and accuracy may be

degraded. A “Shutdown Low Batt” message on the display will appear

for a few seconds before the power is switched off when the battery is

too low to operate the unit.

To extend the battery life:

• Shut down the unit with the power switch when not taking

measurements or use the power save option to have the unit

automatically turn off after 5 minutes.

• Store the unit in a cool dry place.
• Fully charge the battery before storing the unit for extended

periods of time.

• Limit backlight use. The unit consumes 3X normal power with

the backlight on.

Set the backlight time option to 10 seconds, or select “Button Control”

and keep the backlight off.

Battery replacement: The lithium-ion battery used in this unit should

last for many years with normal use. When it no longer powers the unit

long enough to meet testing requirements it will need to be replaced.

Lithium-ion batteries that are properly charged and stored do not usually