LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 69

Advertising
background image

2. Press

ENTER

to select

Select

Measure. Press

to

scroll to

Total Chlorine.

Measure Menu

Turbidity - No Blank
Turbidity - With Blank
Free Chlorine

Total Chlorine

12:00:00 001/500

3. Press

ENTER

to select

Total

Chlorine.

Total Chlorine (L)

Scan Blank

Scan Free Chlorine

12:00:00 001/500

4. Rinse a clean tube (0290)

three times with the sample.

Fill the tube to the 10 mL line

with the sample. Cap the

tube. Dry the tube with a lint-

free cloth.

5. Open the meter lid. Insert

the tube into the chamber.

Align the index line on the

tube with the index arrow on

the meter. Close the lid. Press

ENTER

to scan the blank. The

screen will display

Reading...

for about 1 second and then

return to the

Total Chlorine

menu.

Total Chlorine (L)

Scan Free Chlorine

Scan Total Chlorine

12:00:00 001/500

69

Chlorine Analysis

Advertising
This manual is related to the following products: