Troubleshooting guide – LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 77

Advertising
background image

4. Always insert tube into the meter chamber with the same amount of

pressure and to the same depth.

TROUBLESHOOTING GUIDE

TROUBLESHOOTING

PROBLEM

REASON

SOLUTION

“Blank?”

Sample is reading

lower than the blank.

With samples of very

low concentration

reblank or record as

zero. On samples of

higher concentration

reblank and read again.

Flashing

Low battery. Readings

are reliable.

Charge battery or use

USB wall/computer

charger.

“Low Battery”

Battery voltage is very

low. Readings are not

reliable.

Charge battery or use

USB wall/computer

charger.

“Shut Down Low

Batt” Shut Down

Battery is too low to

operate the unit.

Charge battery or use

USB wall/computer

charger.

“Overrange”

Sample is outside of

acceptable range.

Dilute sample and test

again.

“Error1”

High readings

with 90° and 180°

detectors.

Dilute sample by at

least 50% and retest.

Lost in meter menus Reset to factory

default settings.

Follow Procedure on

page 9 or page 26.

Unusually large

negative or

positive readings

when performing

calibration

Incorrect standards

used to calibrate

meter.

Use fresh 0.0 standard

in clean tube. Reset

meter to factory default

settings. Recalibrate

meter.

Meter is not

responsive to

any button press

(including power

button)

Reset with manual

reset switch.

Use a toothpick to

press switch located in

small recess to the left

of the USB port.

77