6. Press

ENTER

to select

Turbidity Calibration.

Turbidity Calibration

Formazin

Japan Standard

12:00:00 001/500

7. Scroll to the desired

calibration option. Select

a calibration option based

on the composition of the

standards that will be used to

calibrate the meter.

Turbidity Calibration

Formazin

Japan Standard

12:00:00 001/500

NOTE: Stablcal® standards below 50 NTU should not be used to

calibrate the LTC3000we/wi. The diluent has a different refractive index

than traditional formazin standards and will affect the results.

8. Press

ENTER

to save the

selection. The screen will

display

Storing... for about

1 second and return to the

Turbidity Options menu.

Press

EXIT

to return to a

previous menu.

Turbidity Options

Turbidity Units

Turbidity Calibration

12:00:00 001/500

Options/Set Up