LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 54

Advertising
background image

CHLORINE

CALIBRATION

Chlorine Standards

The meter should be calibrated with free chlorine standards. The

calibration should be done with a distilled or deionized water blank and

one chlorine standard of known concentration. The concentration of the

calibration standard should be similar to the expected concentration of

the sample that will be tested. The default reagent system is DPD tablet

reagents.

Chlorine Calibration Procedure

DPD Tablet Reagents

Select Tablet Calibration in the Options and Setup Menu.

1. Press and briefl y hold

to turn the meter on. The

LaMotte logo screen will

appear for about 3 seconds

and the

Main Menu will

appear.

Main Menu

Measure

Data Logging
Options
Run PC Link

12:00:00 001/500

2. Press

ENTER

to select

Measure. Press

to scroll

to

Free Chlorine.

Measure Menu

Turbidity - No Blank
Turbidity - With Blank

Free Chlorine

Total Chlorine

12:00:00 001/500

3. Press

ENTER

to select

Free

Chlorine.

Free Chlorine (T)

Scan Blank

Scan Free Chlorine

12:00:00 001/500

54

Chlorine Calibration