LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 6

Advertising
background image

Accuracy:

Turbidity: From 0-2.5 NTU/FNU the accuracy

is ±0.05 NTU/FNU. From 2.5-100 NTU/FNU

the accuracy is ±2%. Above 100 NTU/FNU the

accuracy is ±3%.

Color: ±15 cu

Chlorine:
TABLET:

0-1.0 ppm Range: ±0.03 ppm

1.0-3.0 ppm Range: ±0.06 ppm

3.0-6.0 ppm Range: ±0.3 ppm

6.0-10.0 ppm Range: ±2.5 ppm
LIQUID:

0-0.5 ppm Range: ±0.03 ppm

0.6-3.0 ppm Range: ±0.06 ppm

3.0-8.0 ppm Range: ±0.4 ppm

8.0-10.0 ppm Range: ±1.5 ppm

Detection Limit:

Turbidity: 0.05 NTU/FNU

Color: 20 cu

Chlorine: 0.03 ppm

Light Source:

Turbidity: Tungsten lamp 2300°C ±50 °C,

LTC3000we; IR LED 850 nm ±10 nm, spectral

bandwidth 50 nm, LTC3000wi

Color: 428 ±2 nm UV LED

Chlorine: 525 ±2 nm LED

Detector

Turbidity: Photodiode, centered at 90°, maximum

peak 400-600 nm, LTC3000we

Photodiode, centered at 90°, LTC3000wi

Color: Photodiode

Chlorine: Photodiode

Response Time:

<2 seconds

Signal

Averaging:

Turbidity

Sample

Chamber:

Accepts 25 mm fl at-bottomed test tubes

Sample:

10 mL in capped tube

Display:

Graphic Liquid Crystal Display

Software:

Auto Shut-off: 5, 10, 30 min, disabled

Calibration: Field adjustable, blank and 1 point

Data Logging: 500 points

Languages:

English, Spanish, French, Portuguese, Italian,

Chinese, Japanese (Kana)

Temperature:

Operation: 0–50 °C; Storage: -40–60 °C

Operation

Humidity Range:

0–90 % RH, non-condensing

6

Advertising
This manual is related to the following products: