LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 45

Advertising
background image

45

19. Press

ENTER

to select

Set

Calibration and save the

calibration. Press

or

to scroll and select

Factory Setting to revert to

the factory calibration. The

meter will momentarily display

Storing... and return to the

Turbidity -With Blank menu.

The calibration has now been

saved and the meter can be

used for testing.

Turbidity WB (F)

Scan Blank

Scan Sample

12:00:00 001/500

NOTE: For the greatest accuracy during the calibration procedure,

be sure that after the meter is blanked and the blank is scanned as a

sample, the reading is 0.00. If not, reblank the meter and scan the blank

again until it reads 0.00. When scanning the calibration standards as

the sample, scan the calibration standard three times removing the

tube from the chamber after each scan and reinserting the tube in the

chamber with the same orientation. The readings should be consistent.

Use the last consistent reading to calibrate the meter. If the readings are

not consistent, avoid using an aberrant reading to calibrate the meter.
ANALYSIS WITHOUT BLANKING PROCEDURE

To obtain the most accurate results the meter should be blanked before

measuring a sample. The blanking step is not as critical for samples

above 10 NTU. The meter should always be blanked before reading

samples below 10 NTU.

1. Press and briefl y hold

to turn the meter on. The

LaMotte logo screen will

appear for about 3 seconds

and the

Main Menu will

appear.

Main Menu

Measure

Data Logging
Options
Run PC Link

12:00:00 001/500

2. Press

ENTER

to select

Measure.

Measure Menu

Turbidity - No Blank

Turbidity - With Blank
Free Chlorine
Total Chlorine

12:00:00 001/500

Turbidity Analysis