LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 73

Advertising
background image

4. Press

ENTER

to select

Color.

Color

Scan Blank

Scan Sample

12:00:00 001/500

5. Rinse a clean tube (0290)

with color-free (distilled or

deionized) water. Fill the tube

to the 10 mL line with the

color-free water. Cap the tube.

Dry the tube with a lint-free

cloth.

6. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid. Press

ENTER

to

scan the blank. The screen

will display

Blank Done for

about 1 second and then

return to the

Color menu.

Color

Scan Blank

Scan Sample

12:00:00 001/500

7. Remove the tube from the

meter. Empty the tube.

Rinse the tube with the color

standard. Fill the tube to the

10 mL line with the color

standard. Cap the tube. Dry

the tube with a lint-free cloth.

8. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid. Press

ENTER

to select

Scan Sample. The

screen will display

Reading...

for about 1 second. The result

will appear on the screen

and then return to the

Color

menu.

Color

110

cu

Scan Blank

Scan Sample

12:00:00 001/500

73

Color Calibration

Advertising
This manual is related to the following products: